Pavić, Arsen and Kodvanj, Janoš and Surjak, Martin (2013) Određivanje stabilnosti novog vanjskog fiksatora primjenom mjernog sustava Aramis. = Determining the stability of novel external fixator by using measuring system Aramis. Tehnički vjesnik : znanstveno-stručni časopis tehničkih fakulteta Sveučilišta u Osijeku, 20 (6). pp. 995-999. ISSN 1330-3651. Vrsta rada: ["eprint_fieldopt_article_type_article" not defined]. . Točan broj autora: 3.
|
Text
Determining the stability of novel external fixator by using measuring system Aramis.pdf - Accepted Version Jezik dokumenta:English Download (1MB) | Preview |
Abstract (Croatian)
Vanjski fiksatori koriste se za liječenje otvorenih i zatvorenih prijeloma dugih kostiju. Krutost i čvrstoća fiksatora uvelike utječu na konačan ishod liječenja. Odgovarajuća stabilnost koštanih ulomaka omogućuje bolje cijeljenje prijeloma i bržu rehabilitaciju pacijenta. Cilj ovog istraživanja bila je procjena biomehaničke stabilnosti novokonstruiranog unilateralnog vanjskog fiksatora za prijelome proksimalne tibije. S ciljem usporedbe svojstava prototipa novog fiksatora s vanjskim fiksatorom Orthofix, koji se često upotrebljava u kliničkoj praksi, provedena su statička i dinamička ispitivanja. Pri statičkom ispitivanju uzorci su opterećeni aksijalno i na savijanje, a pomaci su određeni bezkontaktnim optičkim mjernim sustavom Aramis. Rezultati su pokazali da su biomehanička svojstva nove konstrukcije bolja kod savijanja, dok se konvencionalni vanjski fiksator pokazao boljim pri aksijalnom opterećenju. Pri cikličkom ispitivanju nije bilo značajne razlike u rezultatima.
Abstract
External fixators are used to treat open and closed fractures of long bones. Stiffness and strength of the fixator greatly affect the final outcome of treatment. An adequate stability of bone fragments provides better fracture healing and early rehabilitation of patients. The aim of this study was to evaluate biomechanical stability of the newly developed unilateral external fixator for proximal tibial fractures. In order to compare the properties of the prototype of the novel fixator and commonly used external fixator Orthofix, static and dynamic tests were conducted. Two different types of static tests were carried out, axial compression test and bending test. In static tests displacements were measured and calculated by using the non-contact optical measuring system Aramis. The results indicated that the novel fixator has better outcomes in bending, while the conventional fixator outperformed the new one in axial compression. There was no significant difference in cyclic test.
Item Type: | Article (["eprint_fieldopt_article_type_article" not defined]) |
---|---|
Keywords (Croatian): | external fixator ; optical system Aramis ; stability |
Subjects: | TECHNICAL SCIENCE > Mechanical Engineering |
Divisions: | 200 Department of Engineering Mechanics > 220 Chair of Experimental Mechanics |
Indexed in Web of Science: | Yes |
Indexed in Current Contents: | No |
Citations SCOPUS: | 0 (22.11.2016.) |
Date Deposited: | 22 Apr 2016 10:31 |
Last Modified: | 22 Nov 2016 15:21 |
URI: | http://repozitorij.fsb.hr/id/eprint/5737 |
Actions (login required)
![]() |
View Item |