Interlaboratory comparison in the range from 0 to 2 MPa for Croatian accredited laboratories

Grgec Bermanec, Lovorka and Zvizdić, Davor and Praštalo, Nikola (2012) Interlaboratory comparison in the range from 0 to 2 MPa for Croatian accredited laboratories. = Interlaboratory comparison in the range from 0 to 2 MPa for Croatian accredited laboratories. In: CAFMET 2012 -4th International Metrology Conference, 23.-27.4.2012., Marrakech, Maroko.

Full text not available from this repository.
Official URL: http://bib.irb.hr/prikazi-rad?rad=581318

Abstract

This paper describes procedure and results of the interlaboratory comparison that was organised by the Croatian national pressure laboratory (HMI/FSB-LPM) between eight accredited Croatian pressure calibration laboratories. Comparison was performed in the gauge pressure range from 0 to 2 MPa with pressure transducer as transfer standard and using gas as transmitting medium. Basic aim of this comparison was to improve measurement methods and the measurement uncertainties of each participating laboratory. Analysis of the measurement results will be given by calculating the deviation, En, normalized with respect to the stated measurement uncertainties.

Item Type: Conference or Workshop Item (Lecture)
Keywords (Croatian): interlaboratory comparison; pressure; measurement uncertainty
Subjects: TECHNICAL SCIENCE > Mechanical Engineering
Divisions: 300 Department of Thermodynamics, Thermal and Process Engineering > 320 Chair of Thermal and Process Engineering
Indexed in Web of Science: No
Indexed in Current Contents: No
Date Deposited: 07 Feb 2017 13:48
Last Modified: 07 Feb 2017 13:48
URI: http://repozitorij.fsb.hr/id/eprint/7234

Actions (login required)

View Item View Item

Nema podataka za dohvacanje citata